Do you really need to do it? But why?
Lets start by looking at what can happen underneath the lens of your ultrasound probe. Here is a picture of a phased array ultrasound probe with a tiny little cut in the corner and then the resultant corrosion on the acoustic array because it was not dealt with quickly:
For the full article on this case study, click here
So by ignoring the lens damage, fluid ingress will result which is a massive cross contamination hazard but also will damage the array which will then become a more expensive repair or result in the probe being non economical to repair.
Here is a couple of other examples showing lens damage:
In addition to the obvious cross contamination hazard, there is an electrocution hazard presented by a damaged lens. In this video we show a scenario with a double fault. The lens of the probe has a failure and so does the ultrasound machine. Glad this was in a controlled environment and not on a patient!
For more information from this article, click here
Even sealant missing from around the edge of the probe can cause corrosion to the array as can be seen in this case study
So you’ve been told probes can’t be repaired? We have heard that so many times before. Not true! here is a before/after image of a lens that we recently repaired: